| Does the
Spin Air require calibration? |
| How often
do I need to calibrate the Spin Air? |
| What kind
of battery does the Spin Air use? |
| Nothing
on display. What can I do? |
| What
kinds of bacteria can the Spin Air actually test for? |
|
| What
size of agar plates does the Spin Air use? |
|
| How
can I clean the Spin Air? |
|
| Is
it possible to use disinfectant solution? |
|
| How
does the Spin Air work? |
|
| Can
I set the delay time on the Spin Air? |
|
| What
is the total amount of air that can run thought the
Spin Air? |
|
| What
kind of attachments can be used with the Spin Air? |
|
| What
is the “Slave” unit being used for? |
|
| Do
you Service, support and Train the Spin Air? |
|
| Can
I try out an instrument before I buy it? |
|
| |
| Yes. The Spin Air requires
in house calibration at Neu-Tec facility. The Spin Air
is delivered with a calibration certificate indicating
date of calibration and is valid for one year. |
|
| We recommend calibrating
the Spin Air at least once a year although excessive
use may call for more often calibrations. |
|
| The Spin air uses rechargeable
Ni-Metal Hydride 7.2V internal battery pack, which allows
the Spin Air to work up to 8 hours (without the slave
unit connected). |
|
| Push up button
(↑) for 3 sec. If nothing happens charge batteries
for 4 hours. You can also work with the Spin Air while
it is connected to electricity although it will limit
your range of usage due to requirement of an outlet. |
|
| The spin air
can collect any biological aerosols including molds,
fungi, pollen, spores etc. in accordance to the selectivity
of agar plate that is being used. |
|
| The Spin Air can use both
90mm and 60mm plate according to the attachment that
is being used. |
|
| The Spin Air
may be cleaned with a wet towel with a solution of 80%
alcohol. No residues will remain on the instrument. |
|
| The use of disinfectant is
not recommended due to the fact that they may leave behind
residues. |
|
After placing an agar plate
in the Spin Air and setting the volume of air one would
like to run though it, the Spin Air platform will start
to rotate and spin the agar plate in a set speed (1-4
RPM).
At the same time, air will be sucked in (100 L / min).
This spinning motion provides full coverage of the agar
plate surface and eliminates overlapping layers of bacteria. |
|
| Yes, when needed, the delay
time can be set for up to 1 hour (divided by seconds). |
|
| The Spin Air can be set to
take any amount of air between 0-to-9999 L of air. |
|
| The Spin Air can be connected
to several devices, such as bar code reader, printer
or PC (RS 232) and a Slave unit which allows performing
duplicate experiments or saving time when doing multiple
plates types. |
|
| The “Slave” unit
is a down scale unit that can be connected to the Spin
Air. This unit does not require any setting, it will
simply duplicate the data from the main unit and perform
the same experiment simultaneity resulting accurate
duplicate results. |
|
| Yes, Neu-tec will provide
support and training on the instrument over the phone,
and if needed to will service the instrument at our
laboratory. |
|
| Yes, we rent a wide variety
of instrumentation and offer loaners Call us toll free
1-888-810-5179 for more information. |
|
| |